Notes:
- The RNA Cleanup Binding Buffer may form a precipitate. Should this occur, warm to room temperature to re-dissolve before use.
- Add 4 volumes of molecular biology grade ethanol (≥ 95%) to one volume of RNA Cleanup Washing Buffer.
·
- Centrifugation steps should be at room temperature
·
- Intact Genomics IG®RNA Cleanup Kit will purify RNA ≥ 25 nt. A modification in step 2 allows for the purification of RNA as small as 15 nt (the addition of 2 volumes of ethanol shifts the cutoff size of RNA binding from 25 nt down to 15 nt)
1. Add 100 μl RNA Cleanup Binding Buffer to the 50 μl sample.
– – The starting sample volume of 50 μl is recommended, although for smaller samples, nuclease-free water can be used to adjust the volume. If your sample is larger than 50µl, please scale buffer columns accordingly.
2. Add 150 μl of ethanol (≥ 95%) to your sample and mix by pipetting or flicking the tube. Do not vortex.
–This is for binding RNA ≥ 25 nt in length. Should you need to bind RNA ≥15 nt, add 300 μl of ethanol to your sample instead of 150 μl.
3. Insert the included column into a collection tube and load your sample onto the column. Close the cap and then spin for 1 minute. Discard the flow-through.
–If your sample is diluted > 900 µl, load a portion of the sample. Then spin and repeat as necessary.
4. Insert the column back into the collection tube. Add 500 μl RNA Cleanup Wash Buffer, spin for 1 minute and discard the flow-through. Repeat this step 1X.
5. Transfer the column to an RNase-free 1.5 ml microfuge tube (not included).
–It is important to note that the tip of the column should not come into contact with the flow through. If this occurs, re-spin for 1 minute to ensure traces of salt/ethanol are not carried over to the final sample.
6. Elute RNA with nuclease-free water according to the following table:

If you cannot find the answer to your problem then please contact us or telephone +44 (0)1954 210 200